Precision measurements of quantum defects in the nP3/2 Rydberg States of 85Rb

Abstract

Rydberg States are used in our One Atom Maser experiment because they offer a large dipole moment and couple strongly to low numbers of microwave photons in a high Q cavity. Here we report the absolute frequencies of the P3/2 states for principal quantum numbers n=36 to n=63. These measurements were made with a three step laser excitation scheme. A wavemeter was calibrated against a frequency comb to provide accurate absolute frequency measurements over the entire range, reducing the measurement uncertainty to 1MHz. We compare the spectroscopic results with known frequency measurements as a test of measurement accuracy.

0

Discussion (0)

Sign in to join the discussion.

Loading comments…