A frequency determination method for digitized NMR signals

Abstract

We present a high precision frequency determination method for digitized NMR FID signals. The method employs high precision numerical integration rather than simple summation as in many other techniques. With no independent knowledge of the other parameters of a NMR FID signal (phase φ, amplitude A, and transverse relaxation time T2) this method can determine the signal frequency f0 with a precision of 1/(8π2f02T22) if the observation time T is long enough. The method is especially convenient when the detailed shape of the observed FT NMR spectrum is not well defined. When T2 is +∞ and the signal becomes pure sinusoidal, the precision of the method is 3/(2π2f02T2) which is one order more precise than a typical frequency counter. Analysis of this method shows that the integration reduces the noise by bandwidth narrowing as in a lock-in amplifier, and no extra signal filters are needed. For a pure sinusoidal signal we find from numerical simulations that the noise-induced error in this method reaches the Cramer-Rao Lower Band(CRLB) on frequency determination. For the damped sinusoidal case of most interest, the noise-induced error is found to be within a factor of 2 of CRLB when the measurement time T is a few times larger than T2.We discuss possible improvements for the precision of this method.