K-edge subtraction vs. A-space processing for x-ray imaging of contrast agents: SNR
Abstract
Purpose: To compare two methods that use x-ray spectral information to image externally administered contrast agents: K-edge subtraction and basis-function decomposition (the A-space method), Methods: The K-edge method uses narrow band x-ray spectra with energies infinitesimally below and above the contrast material K-edge energy. The A-space method uses a broad spectrum x-ray tube source and measures the transmitted spectrum with photon counting detectors with pulse height analysis. The methods are compared by their signal to noise ratio (SNR) divided by the patient dose for an imaging task to decide whether contrast material is present in a soft tissue background. The performance with iodine or gadolinium containing contrast material is evaluated as a function of object thickness and the x-ray tube voltage of the A-space method. Results: For a tube voltages above 60 kV and soft tissue thicknesses from 5 to 25 g/cm2, the A-space method has a larger SNR per dose than the K-edge subtraction method for either iodine or gadolinium containing contrast agent. Conclusion: Even with the unrealistic spectra assumed for the K-edge method, the A-space method has a substantially larger SNR per patient dose.