Direct current conditioning to reduce the electrical impedance of the electrode to skin contact in physiological recording and stimulation

Abstract

Direct and alternating current iontophoresis and electro-osmosis methodologies have provided new methods of transcutaneous drug delivery. A byproduct of such methods is lowering the electrical impedance of the electrode to skin contact, as conductive ions permeate the stratum corneum, the primary resistive layer of the skin. We developed a method for adapting iontophoresis to condition the electrode to skin contact, both for electrophysiological recording and electrical stimulation of body tissues. By utilizing direct current to treat electrodes with high impedance we show the effectiveness of iontopheresis as a driving force for permeation of ionic electrolyte into the skin barrier. We applied direct current (DC) levels of 50 μA to electrodes on the human head for 30 seconds with paste (Nihon Kohden Elefix) electrolyte. Typically immediately after DC treatment conditioning there was an impedance drop of 10-30%. The effect was lasting over several hours, with the paste electrolyte. These results demonstrate the feasibility of DC conditioning to reduce the set time of electrolytic solutions and to maintain good skin contact during extended recording or stimulation sessions.

0

Turn this paper into a lesson

ArcXiv compiles a structured reading guide from this paper's metadata: plain-English importance, contributions, prerequisite concepts, which sections to read first, flashcards, and a quiz. Grounded in the abstract, never invented.

Discussion (0)

Sign in to join the discussion.

Loading comments…