Demonstration of resolving power λ/λ > 10,000 for a space-based x-ray transmission grating spectrometer

Abstract

We present measurements of the resolving power of a soft x-ray spectrometer consisting of 200 nm-period lightweight, alignment-insensitive critical-angle transmission (CAT) gratings and a lightweight slumped-glass Wolter-I focusing mirror pair. We measure and model contributions from source, mirrors, detector pixel size, and grating period variation to the natural line width spectrum of the Al and Mg Kα1 α2 doublets. Measuring up to 18 th diffraction order at characteristic Al-K wavelengths we consistently obtain small broadening due to gratings corresponding to a minimum effective grating resolving power Rg > 10,000 with 90\% confidence. Upper limits are often compatible with Rg = ∞. Independent fitting of different diffraction orders, as well as ensemble fitting of multiple orders at multiple wavelengths, gives compatible results. Our data leads to uncertainties for the Al-Kα doublet line width and line separation parameters 2-3 times smaller than values found in the literature. Data from three different gratings are mutually compatible. This demonstrates that CAT gratings perform in excess of the requirements for the Arcus Explorer mission and are suitable for next-generation space-based x-ray spectrometer designs with resolving power 5-10 times higher than the transmission grating spectrometer on the Chandra X-ray Observatory.

0

Turn this paper into a lesson

ArcXiv compiles a structured reading guide from this paper's metadata: plain-English importance, contributions, prerequisite concepts, which sections to read first, flashcards, and a quiz. Grounded in the abstract, never invented.

Discussion (0)

Sign in to join the discussion.

Loading comments…