Evaluation of the blackbody radiation shift of an Yb optical lattice clock at KRISS
Abstract
As optical clocks are improved to reach the frequency uncertainty below the 10-17 level, the frequency shift due to the blackbody radiation (BBR) has been one of the major systematic effects hindering further improvement. To evaluate the BBR shift of an Yb optical lattice clock at KRISS, we installed an in-vacuum BBR shield and made radiation thermometry using a black-coated-sphere thermal probe. After we quantitatively measured the conduction loss of the thermal probe and the effects of all the external radiation sources, we determined the temperature at the atom trap site with an uncertainty of 13 mK, which corresponds to an uncertainty of 0.22 mHz in the clock frequency (a fractional frequency of 4.2×10-19). The total uncertainty of the BBR shift including the atomic response is 9.5×10-19.