SAM: Semi-Active Mechanism for Extensible Continuum Manipulator and Real-time Hysteresis Compensation Control Algorithm

Abstract

Cable-Driven Continuum Manipulators (CDCMs) enable scar-free procedures but face limitations in workspace and control accuracy due to hysteresis. We introduce an extensible CDCM with a Semi-active Mechanism (SAM) and develop a real-time hysteresis compensation control algorithm using a Temporal Convolutional Network (TCN) based on data collected from fiducial markers and RGBD sensing. Performance validation shows the proposed controller significantly reduces hysteresis by up to 69.5% in random trajectory tracking test and approximately 26% in the box pointing task. The SAM mechanism enables access to various lesions without damaging surrounding tissues. The proposed controller with TCN-based compensation effectively predicts hysteresis behavior and minimizes position and joint angle errors in real-time, which has the potential to enhance surgical task performance.

0

Turn this paper into a full lesson

ArcXiv compiles a staged curriculum from this paper: 8-12 lessons across beginner → advanced, synthesised section guides, visuals, flashcards, a quiz, exercises, and on-demand deep dives per section. Grounded in the abstract, never invented.

Discussion (0)

Sign in to join the discussion.