First ex-vivo positronium imaging of tissues with modular J-PET scanner using 44Sc radionuclide

Abstract

This study presents the first ex-vivo positronium imaging of human tissues using the modular J-PET scanner with the 44Sc radionuclide. The 44Sc isotope was produced via the 44Ca(p, n)44Sc nuclear reaction and used to perform positronium imaging of phantom composed of human adipose tissue, cardiac myxoma tissue, thrombi blood clot, and also porous polymer XAD4, and a certified reference material (CRM) made from fused silica. The experiment demonstrates the suitability of 44Sc as a positron source for positronium imaging. The performance of J-PET for positronium imaging with 44Sc was validated by proper reconstruction of the mean orthopositronium lifetime for CRM material and XAD-4 polymer. The mean ortho-positronium (oPs) lifetimes determined for adipose tissue, cardiac myxoma tissues and thrombi were consistent with results of previous experiments. The study highlights the potential 44Sc radionuclide for positronium lifetime imaging (PLI).