A Practical Noise2Noise Denoising Pipeline for High-Throughput Raman Spectroscopy

Abstract

A lightweight and reproducible denoising pipeline for high-throughput Raman spectroscopy is presented. The approach relies on a one-dimensional convolutional autoencoder trained using a Noise2Noise strategy, requiring neither external spectral libraries nor high signal-to-noise reference spectra for training. From a reduced training subset composed of repeated short-exposure acquisitions, the model learns to reconstruct Raman spectra while efficiently suppressing stochastic noise. The method is evaluated on a heterogeneous mineral sample, using both quantitative spectral fidelity metrics (RMSE, SNR, SSIM) and task-oriented criteria based on unsupervised K-means classification. Results demonstrate that integration times as short as 5 ms per spectrum, which are typically insufficient for reliable interpretation, yield denoised spectra with high fidelity to the reference data while preserving chemically coherent maps. This work provides a practical trade-off between spectral quality and acquisition speed, enabling fast, adaptable Raman workflows compatible with routine laboratory use. It also offers a transferable framework for other one-dimensional spectroscopic modalities.

0

Turn this paper into a full lesson

ArcXiv compiles a staged curriculum from this paper: 8-12 lessons across beginner → advanced, synthesised section guides, visuals, flashcards, a quiz, exercises, and on-demand deep dives per section. Grounded in the abstract, never invented.

Discussion (0)

Sign in to join the discussion.

Loading comments…