Evaluation of a long-time temperature drift in a commercial Quantum Design MPMS SQUID magnetometer using Gd2O3 as a standard
Abstract
The long-time temperature drift in a commercial Quantum Design MPMS SQUID magnetometer was evaluated using time-dependent magnetization measurements of Gd2O3. In contrast to earlier claims, the amplitude of the drift was found not to exceed 1-1.5 K. 30 minutes after system stabilization the temperature deviation did not exceed 0.2 K and the temperature was fully stabilized in less than 3 hours.