Evaluation of a long-time temperature drift in a commercial Quantum Design MPMS SQUID magnetometer using Gd2O3 as a standard

Abstract

The long-time temperature drift in a commercial Quantum Design MPMS SQUID magnetometer was evaluated using time-dependent magnetization measurements of Gd2O3. In contrast to earlier claims, the amplitude of the drift was found not to exceed 1-1.5 K. 30 minutes after system stabilization the temperature deviation did not exceed 0.2 K and the temperature was fully stabilized in less than 3 hours.

0

Turn this paper into a full lesson

ArcXiv compiles a staged curriculum from this paper: 8-12 lessons across beginner → advanced, synthesised section guides, visuals, flashcards, a quiz, exercises, and on-demand deep dives per section. Grounded in the abstract, never invented.

Discussion (0)

Sign in to join the discussion.

Loading comments…