Precise On-line Position Measurement for Particle Therapy

Abstract

An on-line beam position monitoring and regular beam stability tests are of utmost importance for the Quality Assurance (QA) of the patient treatment at any particle therapy facility. The Gantry0.5 mm2 at the Paul Scherrer Institute uses a strip ionization chamber for the on-line beam position verification. The design of the strip chamber placed in the beam in front of the patient allows for a small beam penumbra in order to achieve a high-quality lateral beam delivery. The position error of 1 mm in a lateral plane (plane perpendicular to the beam direction) can result in a dose inhomogeneity of more than 5 \%. Therefore the goal of Gantry 2 commissioning was to reach a sub-millimeter level of the reconstruction accuracy in order to bring a dose uncertainty to a level of 1 \%. In fact, we observed that for beams offered by Gantry 2 signal profiles in a lateral plane can be reconstructed with a precision of 0.1 mm. This is a necessary criterion to perform a reliable patient treatment. The front end electronics and the whole data processing sequence have been optimized for minimizing the dead time in between two consecutive spots to about 2 ms: the charge collection is performed in about 1 ms, read-out takes place in about 100 μs while data verification and logging are completed in less than 1 ms.