Waveform Timing Performance of a 5 GS/s Fast Pulse Sampling Module with DRS4
Abstract
We first clarify timing issues of non-uniform sampling intervals regarding a 5 GS/s fast pulse sampling module with DRS4. Calibration strategy is proposed, and as a result, the waveform timing performance is improved to be below 10 ps RMS. We then further evaluate waveform timing performance of the module by comparing with a 10 GS/s oscilloscope in a setup with plastic scintillators and fast PMTs. Different waveform timing algorithms are employed for analysis, and the module shows comparable timing performance with that of the oscilloscope.