Evaluation of a front-end ASIC for the readout of PMTs in large dynamic range

Abstract

The Large High Altitude Air Shower Observatory (LHAASO) project has been proposed for the survey and study of cosmic rays. In the LHAASO project, the Water Cherenkov Detector Array (WCDA) is one of major detectors for searching gamma ray sources. A Charge-to-Time Convertor (QTC) ASIC (Application Specification Integrated Circuit) fabricated in Global Foundry 0.35 μm CMOS technology, has been developed for readout of Photomultiplier Tubes (PMTs) in the WCDA. This paper focuses on the evaluation of this front-end readout ASIC performance. Test results indicate that the time resolution is better than 400 ps and the charge resolution is better than 1% with large input signals and remains better than 15% @ 1 Photo Electron (P.E.), both beyond the application requirement. Moreover, this ASIC has a weak ambient temperature dependence, low input rate dependence and high channel-to-channel isolation.

0

Discussion (0)

Sign in to join the discussion.

Loading comments…