Measurement of the response of a liquid scintillation detector to monoenergetic electrons and neutrons

Abstract

The response of the liquid scintillator (EJ-301 equivalent to NE-213) to the monoenergetic electrons produced in Compton scattered γ-ray tagging has been carried out for various radioactive γ-ray sources. The measured electron response is found to be linear up to 4~MeVee and the resolution of the liquid scintillator at 1~MeVee is observed to be ~11\%. The pulse shape discrimination and pulse height response of the liquid scintillator for neutrons has been measured using 7Li(p,n1)7Be*(0.429 MeV) reaction. Non linear response to mono-energetic neutrons for the liquid scintillator is observed at En=5.3, 9.0 and 12.7 MeV. The measured response of the liquid scintillator for electrons and neutrons have been compared with Geant4 simulation.