Initial Measurements with the PETsys TOFPET2 ASIC Evaluation Kit and a Characterization of the ASIC TDC
Abstract
For a first characterization, we used the two KETEK-PM3325-WB SiPMs each equipped with a 3x3x5 mm3 LYSO scintillation crystal provided with the PETsys TOFPET2 ASIC Evaluation Kit. We changed the lower of two discriminator thresholds (DT1) in the timing branch from vtht1 = 5 - 30. The overvoltage was varied in a range of 1.25 V - 7.25 V. The ambient temperature was kept at 16C. For all measurements, we performed an energy calibration including a correction for saturation. We evaluated the energy resolution, the coincidence resolving time (CRT) and the coincidence rate. At an overvoltage of 6 V, we obtained an energy resolution of about 10% FWHM, a CRT of approximately 210 ps FWHM and 400 ps FWTM, the coincidence rate showed only small variations of about 5%. To investigate the influence of the ambient temperature, it was varied between 12C - 20C. At 12C and an overvoltage of 6.5 V, a CRT of approx. 195 ps FWHM and an energy resolution of about 9.5% FWHM could be measured. Observed satellite peaks in the time difference spectra were investigated in more detail. We could show that the location of the satellite peaks is correlated with a programmable delay element in the trigger circuit.