Development of an "α-ToF" detector for correlated measurement of atomic masses and decay properties

Abstract

We have developed a novel detector, referred to as an "α-ToF detector", for correlated measurements of atomic masses and decay properties of low-yield, short-lived radioactive isotopes using a multi-reflection time-of-flight mass spectrograph. By correlating measured time-of-flight signals with decay events, it will be possible to suppress background events and obtain accurate, high-precision mass values even in cases of very low event rates. An offline test of the α-ToF detector has shown that the time-of-flight detection efficiency for 5.48~MeV α-rays is more than 90\% and yields a time resolution of 251.5(68)~ps and an energy resolution of 141.1(9)~keV. Using a two-dimensional spectrum of the correlated α-ray energy and time-of-flight, the α-rays from mixed α sources could be fairly well resolved.

0

Turn this paper into a lesson

ArcXiv compiles a structured reading guide from this paper's metadata: plain-English importance, contributions, prerequisite concepts, which sections to read first, flashcards, and a quiz. Grounded in the abstract, never invented.

Discussion (0)

Sign in to join the discussion.

Loading comments…