Analysis and mitigation of pulse-pile-up tail artifacts in warm-plasma pulse-height X-ray spectra

Abstract

Pulse pile-up in pulse-height energy analyzers increases when the incident rate of pulses increases relative to the inverse of the dead time per pulse of the detection system. Changes in the observed energy distributions with incident rate and detector-electronics-formed pulse shape then occur. We focus on weak high energy tails in X-ray spectra, important for measurements on partially ionized, warm, pure-hydrogen plasma. A first-principles two-photon pulse-pile-up model is derived specific to trapezoidal-shaped pulses; quantitative agreement is found between the measurements and the model predictions. The modeling is then used to diagnose pulse-pile-up tail artifacts and mitigate them in relatively low count-rate spectra.