Optical-pump terahertz-probe spectroscopy in high magnetic fields with kHz single-shot detection
Abstract
We demonstrate optical pump/THz probe (OPTP) spectroscopy with a variable external magnetic field (0-9 T) in which the time-dependent THz signal is measured by echelon-based single-shot detection at a 1 kHz repetition rate. The method reduces data acquisition times by more than an order of magnitude compared to conventional electro-optic sampling using a scanning delay stage. The approach illustrates the wide applicability of the single-shot measurement approach to nonequilibrium systems that are studied through OPTP spectroscopy, especially in cases where parameters such as magnetic field strength (B) or other experimental parameters are varied. We demonstrate the capabilities of our measurement by performing cyclotron resonance experiments in bulk silicon, where we observe B-field dependent carrier relaxation and distinct relaxation rates for different carrier types. We use a pair of economical linear array detectors to measure 500 time points on each shot, offering equivalent performance to camera-based detection with possibilities for higher repetition rates.
Turn this paper into a lesson
ArcXiv compiles a structured reading guide from this paper's metadata: plain-English importance, contributions, prerequisite concepts, which sections to read first, flashcards, and a quiz. Grounded in the abstract, never invented.