First (calibration) experiment using proton beam from FRENA at SINP
Abstract
This work presents the first calibration experiment of a 3 MV Tandetron accelerator, FRENA, performed in May 2022. The 7Li(p,n) reaction threshold was measured to calibrate the terminal voltage measuring device. A LiF target of thickness 175 μg/cm2 was used in the experiment. The measured threshold was 18722.7 keV, indicating 6-10 keV energy shift.