Comparative study of YSO, GaGG, and BGO scintillators coupled to a SiPM array for gamma-ray spectroscopy
Abstract
A compact gamma-ray detector module was developed and characterized for spectroscopic applications across a wide energy range. The detector comprises a 4x4 array of silicon photomultipliers (SiPMs, Hamamatsu MPPC S13360-3050) coupled with three different inorganic scintillators (YSO, GaGG, BGO). The performance of the detector was evaluated over an energy range from 32 keV to 2500 keV using various standard gamma-ray point sources (241Am, 137Cs, 22Na, 60Co, 228Th). All three detector configurations demonstrated linearity between pulse height and gamma-ray energy over most of the tested range. GAGG exhibited the most consistent linear behavior up to 2500 keV. YSO with a faster decay constant showed linearity up to ~2000 keV, followed by slight saturation effects likely due to SiPM microcell occupancy limits at higher photon flux. BGO, characterized by a low light output and longer decay time, demonstrated stable linearity up to 2500 keV but reduced sensitivity below 300 keV. The measured energy resolution at 662 keV was 8.17% for GAGG, 9.3% for YSO, and 10.2% for BGO, consistent with their light yield and photon detection efficiency (PDE) - matching characteristics. The obtained results showed the trade-offs between scintillator materials in terms of energy resolution, dynamic range, and linearity, and confirm the suitability of the tested configurations for compact gamma-ray spectroscopy applications across the studied energy range.
Turn this paper into a full lesson
ArcXiv compiles a staged curriculum from this paper: 8-12 lessons across beginner → advanced, synthesised section guides, visuals, flashcards, a quiz, exercises, and on-demand deep dives per section. Grounded in the abstract, never invented.