Modelling the Performance of Tritium Process Monitors from First Principles

Abstract

Ionization chamber-based, in-line tritium process monitors play an important part in determining the behavior of a tritium system. The one-liter detection volume monitor has been characterized well through experiment and to respond linearly to tritium concentrations for the range of 1 μCi/m3 to 1 Ci/m3. Additionally, it has been shown to behave nonlinearly for low voltage on the central anode and low pressure of the carrier gas. A computational model was developed from first principles that successfully describes the behavior of the one-liter monitor for each of these regimes. Predictions from the model are compared to previously-collected experimental data in order to determine validity and tune the model. The model will be expanded to incorporate additional detector geometries and designs in the future.

0

Turn this paper into a full lesson

ArcXiv compiles a staged curriculum from this paper: 8-12 lessons across beginner → advanced, synthesised section guides, visuals, flashcards, a quiz, exercises, and on-demand deep dives per section. Grounded in the abstract, never invented.

Discussion (0)

Sign in to join the discussion.

Loading comments…