Picosecond laser test unit for photosensor characterization at ambient and low temperatures
Abstract
Accurate single photoelectron (SPE) characterization of photosensors is essential for controlling systematic uncertainties in low-light neutrino and dark matter detectors. We present a compact laboratory setup for the characterization of photosensors under controlled, low-light conditions. Specifically, we demonstrate its use with photomultiplier tubes (PMTs) operated at the SPE-level, using picosecond laser pulses and waveform digitization to determine key PMT properties. Measurements as a function of supply voltage and temperature (-50 to +20) are performed on ET Enterprises 9821(Q)B tubes and a Hamamatsu R9980 assembly, which show exponential gain-voltage behavior and device-to-device variation. Cooling increases the gain by 0.1\,\%/, while the transit time spread (TTS) and peak-to-valley ratio (P/V) exhibit no clear temperature dependence. TTS decreases with voltage. Late pulses remain at the percent level and prepulses at the sub-percent level. Cable length affects both apparent gain and TTS. A model-independent, data-driven self-convolution method is introduced to quantify double photoelectron contributions from pulse charge spectra. The procedures provide a reproducible, practice-oriented reference for SPE-level PMT characterization and can be extended to other photosensor types.