SiPM non-linearity studies in beam tests with scintillating crystals
Abstract
High-granularity homogeneous electromagnetic calorimeters based on scintillating crystals and silicon photomultipliers (SiPMs) are a promising option for future e+e- Higgs factories, where both excellent energy resolution and a very large dynamic range are required. In this work, the non-linear response of high-pixel-density SiPMs with pixel pitches of 6--10~μm coupled to BGO and BSO crystals is studied under realistic beam conditions. A dual-end readout scheme with an attenuated reference SiPM was employed to precisely calibrate the deposited energy and the corresponding number of photoelectrons over a wide dynamic range. Beam tests were carried out at the CERN SPS H2 beamline using high-energy electrons, with a tungsten pre-shower and variable incident angles to enhance energy deposition. The measurements directly quantify the non-linear response of SiPMs to scintillation light over an extended dynamic range. For BGO-coupled Hamamatsu SiPMs, deviations from linearity of about 20\% are observed at 5×105 photoelectrons, while larger deviations are measured for the tested NDL devices and for configurations with faster BSO scintillation.