Thermal response of an in-situ STEM MEMS chip under rapid pulse heating

Phillip Dumitraschkewitz, Thomas Kremmer

cond-mat.mtrl-sciarXiv:2609.08473

Abstract

In-situ rapid solidification studies demand measurements of thermal histories with high temporal resolution. We present a simple, effective setup to quantify the cooling response of an uncoated commercial Protochips Fusion MEMS chip in an in-situ scanning transmission electron microscopy (STEM) context. We drive user-defined temperature programs via an arbitrary waveform generator (AWG), while recording the voltage drops across a series shunt to reconstruct chip resistance and temperature at sub-millisecond resolution. We confirm the response times inferred from the current; however, the temperature obtained from the physically linked resistance, T(R), evolves more slowly. Analysis of the maximum cooling step reveals an exponential-like relaxation with time constant τ=1.80 ms, consistent with reported thermal lag constants for fast scanning calorimetry. From the time to reach 95\% of the temperature difference ΔT, we measure an average cooling rate of ≈ 7.9× 104 K/s. Robustness checks include repeated R(T) measurements (revealing a modest downward drift approaching an asymptote), a 10 kΩ test load, and characterization of small off-duty arbitrary waveform generator leakage/offsets. These findings define practical bounds on achievable thermal-path rates when planning in-situ electron microscopy experiments with this chip platform.