Towards in vivo photon dosimetry using radiation-sensitive mixed-core ultrasound phase-change contrast agents
Bram Carlier, Reza Bahrami Gorji, Qingyuan Tan, Meritxell Gambús-Marfà, Uwe Himmelreich, Meng-Xing Tang, Koen Van Den Abeele, Sophie V. Heymans
Abstract
Objective. Advanced radiotherapy applies complex treatment plans that require verification of the beam delivery. Therefore, in vivo dosimetry, measuring delivered radiation doses directly on the patient, has become increasingly important. Radiation-sensitive phase-change nanodroplets were recently proposed for in vivo radiation sensing. The radiation-induced conversion of superheated nanodroplets into microbubbles generates ultrasound contrast from which radiation deposition can be derived. While successful for in vivo proton range verification, previously used nanodroplet formulations were not suitable for in vivo photon dosimetry due to stringent requirements on the degree of superheat of the nanodroplet core. Here, we used a mixed-core strategy, combining different perfluorocarbons in the same nanodroplet core, to modulate the photon response and droplet stability at body temperature. Approach. Lipid-shelled nanodroplets with different cores (pure octafluoropropane, pure decafluorobutane and a 17:4 (v/v) octafluoropropane:decafluorobutane mixture) were dispersed in deionized water in phantom containers and irradiated with a photons at various temperatures. Simultaneous high-frame rate ultrasound acquisition allowed to link ultrasound contrast generation to the radiation delivery. Nanodroplets were characterized using 19F NMR spectroscopy. Main results. Photons induced nanodroplet-mediated ultrasound contrast strong enough for clinical applications at body temperature. A clear trade-off was observed between a stronger photon response and maintaining sufficient droplet stability. Neither octafluoropropane (lacking stability) nor decafluorobutane (lacking photon response) cores proved suitable. On the contrary, the mixed-core formulation exhibited a strong photon response with minimal background already at small radiation doses (0.1 Gy). This paves the way for future in vivo dosimetry.
Create a lesson
Related papers
Physics-Based Differentiable Beamforming for Deep Learning-Based Contrast Enhanced Ultrasound Super-Resolution Imaging
Shared title terminology · Anass Hameddine, Bram de Wilde, Michel Versluis et al.
PET Attenuation Map Generation for Brain Imaging from Virtual Non-Contrast and Virtual Non-Iodine Images Obtained with Photon Counting CT Scanners
Shared title terminology · Heba Alrakh, Lutz Tellmann, Omid Nikoubashman et al.
Sensitivity and Identifiability of Tumor-Related Changes in Longitudinal Microwave Breast Imaging
Shared title terminology · Somayeh B. Shafiei, Ian Jeffrey, Vahab Khoshdel et al.
A single radiation hard pixel detector for proton dosimetry, imaging, energy resolution and beam monitoring
Shared title terminology · Nico Frederic Brosda, Stephane Higueret, Nicolas Arbor et al.
Towards 3D fully randomized frequency-domain reconstruction of the speed of sound in breast ultrasound computed tomography
Shared title terminology · Luca A. Forte
Optical Scintillation Imaging for Dynamic Source Tracking Dosimetry During Prostate HDR Brachytherapy
Shared title terminology · Kevin Willy, Megan Clark, Petr Bruza et al.