Multi-Sensor Fusion for False-Alarm Reduction by Timed Electromechanical Corroboration

Zixuan Fu, Anthony Raffaele, Unnikrishna Pillai

eess.SP

Abstract

A physician checks the clinical electrocardiogram against other observations. A wearable monitor has no physician. This paper shows that a mechanical sensor worn beside the electrocardiogram can take over that check, and that the right added sensors lower false alarms further. A beat is accepted only when a mechanical contraction follows at the delay learned for the wearer. Beat-shaped artifacts injected into resting and exercise recordings passed the electrocardiogram alone almost every time. A seismocardiogram or phonocardiogram as the second channel rejected most while keeping nearly all true beats, strongly at rest and partly during exercise. Shifting the mechanical signal in time removed the benefit. A detection criterion scores an added sensor by the part of its response that the sensors in place cannot predict. With four sensors, this score ranked the sets as their measured false-alarm rates did. With the two sensors it selected, 6.2 percent of the artifacts passed while 95 percent of true beats were kept, half the rate of the best single sensor. In integer arithmetic on a low-cost microcontroller, the detector made the same decisions as the software. These rates come from injected artifacts, not alarms in patients. Natural motion and clinical events remain untested.