Diffusion Posterior Sampling for Mitigation of Spatially-Variant System Blur in Ultra-High-Resolution CT of Trabecular Bone

T. Ye, G. Shi, A. Sivakumar, F. J. Quevedo Gonzalez, K. F. Moseley, J. A. Carrino, J. H. Siewerdsen, W. Zbijewski

physics.med-ph

Abstract

Objective: CT spatial resolution degrades with distance from isocenter. We evaluate diffusion posterior sampling (DPS) deblurring for restoring structural detail and harmonizing texture features (TFs) of off-center trabecular bone to isocenter in ultra-high resolution CT (UHR CT). Methods: DPS pairs a diffusion prior, trained on isocenter UHR CT patches of trabecular bone from four human femora, with a likelihood term enforcing consistency with the observed image. Two variants were tested on a held-out femur at radial shifts of 90, 130 and 160 mm: KnownBlurDPS, with an anisotropic Gaussian kernel derived from modulation transfer function measurements, and BlindDPS, which estimated the kernel jointly with the image. Restored patches were compared with isocenter ground truth by structural similarity (SSIM), gradient magnitude similarity deviation (GMSD), and concordance correlation coefficient (CCC) of 40 gray-level co-occurrence and run-length TFs. Results: Both DPS variants improved SSIM and GMSD by 1-10% and mean TF CCC by 30-130%. The number of reproducible TFs (CCC > 0.80) at 90/130/160 mm rose from 13/2/1 without deblurring to 33/35/29 (KnownBlurDPS) and 40/36/16 (BlindDPS). At 90 mm, the accuracy of a linear classifier predicting the shift from TFs dropped from 79% to chance level with BlindDPS. Conclusion: DPS partly restores trabecular detail and TF reproducibility lost to spatially-variant blur in UHR CT; the blind variant matches the measured-kernel variant up to 130 mm. Significance: DPS deblurring can standardize trabecular bone TFs against patient mis-centering, potentially improving the robustness of TF-based CT radiomics; its blind form needs no CT system characterization.

Create a lesson