Longitudinal-dither wavefront sensing: Calibration-free measurement of mode mismatch

Masayuki Nakano

physics.optics

Abstract

Mode mismatch between a gravitational-wave detector and the beams injected into it lowers the power stored in the arms and the level of squeezing. Correcting it requires the complex amplitudes of the three second-order modes, measured on a calibrated scale with the cavity on resonance at its operating power. Here we present a method that obtains a signal proportional to the mismatch by moving the beam on the wavefront sensors already used for alignment control. A longitudinal dither of the cavity adds to the resonant mode a reference field, which reaches the sensors through the same signal chains and so calibrates both the scale of this signal and the Gouy phases of the sensors and actuators, with no external calibration. We demonstrate the method on an optical cavity held on resonance in its operating state at input powers of 2 W and 62 W. At 2 W it yields an optical model that agrees with independent measurements in the mode content, and with the layout model in the actuator Gouy phase within its uncertainty of about 1∘1^\circ. At 62 W the thermally induced change of the eigenmode is decomposed into waist-size and waist-position changes. The method requires only components that every controlled cavity already possesses and enables continuous in situ monitoring of the mode matching in the operating state.