An x-ray detector using PIN photodiodes for the axion helioscope

Abstract

An x-ray detector for a solar axion search was developed. The detector is operated at 60K in a cryostat of a superconducting magnet. Special care was paid to microphonic noise immunity and mechanical structure against thermal contraction. The detector consists of an array of PIN photodiodes and tailor made preamplifiers. The size of each PIN photodiode is 11× 11× 0.5\ mm3 and 16 pieces are used for the detector. The detector consists of two parts, the front-end part being operated at a temperature of 60K and the main part in room temperature. Under these circumstances, the detector achieved 1.0 keV resolution in FWHM, 2.5 keV threshold and 6× 10-5 counts sec-1 keV-1 cm-2 background level.