Crosstalk Correction in Atomic Force Microscopy
A. Hoffmann, T. Jungk, E. Soergel
Abstract
Commercial atomic force microscopes usually use a four-segmented photodiode to detect the motion of the cantilever via laser beam deflection. This read-out technique enables to measure bending and torsion of the cantilever separately. A slight angle between the orientation of the photodiode and the plane of the readout beam, however, causes false signals in both readout channels, so-called crosstalk, that may lead to misinterpretation of the acquired data. We demonstrate this fault with images recorded in contact mode on ferroelectric crystals and present an electronic circuit to compensate for it, thereby enabling crosstalk-free imaging.