GePEToS : A Geant4 Monte Carlo simulation package for Positron Emission Tomography
S. Jan, J. Collot, M. -L. Gallin-Martel, P. Martin, F. Mayet, E. Tournefier
Abstract
GePEToS is a simulation framework developed over the last few years for assessing the instrumental performance of future PET scanners. It is based on Geant4, written in Object-Oriented C++ and runs on Linux platforms. The validity of GePEToS has been tested on the well-known Siemens ECAT EXACT HR+ camera. The results of two application examples are presented : the design optimization of a liquid Xe micro-PET camera dedicated to small animal imaging as well as the evaluation of the effect of a strong axial magnetic field on the image resolution of a Concorde P4 micro-PET camera.