Development and operation of a pixel segmented liquid-filled linear array for radiotherapy quality assurance
J. Pardo, L. Franco, F. Gomez, A. Iglesias, A. Pazos, J. Pena, R. Lobato, J. Mosquera, M. Pombar, J. Sendon
Abstract
A liquid isooctane (C8H18) filled ionization linear array for radiotherapy quality assurance has been designed, built and tested. The detector consists of 128 pixels, each of them with an area of 1.7 mm × 1.7 mm and a gap of 0.5 mm. The small pixel size makes the detector ideal for high gradient beam profiles like those present in Intensity Modulated Radiation Therapy (IMRT) and radiosurgery. As read-out electronics we use the X-Ray Data Acquisition System (XDAS) with the Xchip developed by the CCLRC. Studies concerning the collection efficiency dependence on the polarization voltage and on the dose rate have been made in order to optimize the device operation. In the first tests we have studied dose rate and energy dependences, and signal reproducibility. Dose rate dependence was found lower than 2.5 % up to 5 Gy min-1, and energy dependence lower than 2.1 % up to 20 cm depth in solid water. Output factors and penumbras for several rectangular fields have been measured with the linear array and were compared with the results obtained with a 0.125 cm3 air ionization chamber and radiographic film, respectively. Finally, we have acquired profiles for an IMRT field and for a virtual wedge. These profiles have also been compared with radiographic film measurements. All the comparisons show a good correspondence. Signal reproducibility was within a 2% during the test period (around three months). The device has proved its capability to verify on-line therapy beams with good spatial resolution and signal to noise ratio.
Create a lesson
Related papers
T2 estimation from MOLLI acquisitions
Catarina N Carvalho, Andreia S Gaspar, Rita G Nunes et al.
The Magnetic-Resonance-Based, Omni-Directional Hydrophone
Nathan Thyberg, Davi Cavinatto, Katia Oler et al.
Closing the gap: A two-site MLC matching study for Varian TrueBeam Linear Accelerators
Michael Douglass, Corey Bridger, Mitchell Herrick et al.
Geant4 Simulation of an X-ray Beam Monitoring Detector based on plastic scintillators
C. H. Zepeda Fernández, E. Moreno Barbosa, J. M. Arredondo-Velázquez et al.
Dosimetric equivalence of deep learning prostate contours after LDR brachytherapy: pre-declared margins, patient-level acceptance thresholds and the incremental predictive value of DVH indices
Louis-Bernard St-Cyr, Anne Saint-Laurent, José Angel Lesteiro-Tejeda et al.
Fail-closed conformal multiresolution error control for conservative three-dimensional dose remapping in synthetic phantoms
Yuntao Wang