Testing and design of a lens system for atom trapping and fluorescence detection

Abstract

We present methods and results of the testing of an inexpensive home-made diffraction limited lens system, the design of which was proposed in a recent paper and which has since been used (with slight alterations) by several research groups. Our system will be used for both: focussing a collimated laser beam at a wavelength of lambda=830 nm down to a narrow spot and for collimating fluorescence light (lambda=780 nm) emitted from rubidium atoms captured in this spot. Useful tests for lens systems include the use of ray tracing software, shear-wave interferometers, the imaging of test charts and of polystyrene beads of a very small size. We present these methods and show how conclusions can be drawn for the design under test.