Inverse identification of thermal-radiative parameters for reduced effective-temperature uncertainty in an ion optical clock

Jie Zhao, Yani Zuo, Shaoyang Dai, Chaowei Wang, Ang Huang, Jian Song, Guangyao Wang, Fangfei Wu, Haotian Shi, Zhengyong Li, Fang Fang

physics.atom-pharXiv:2610.14796

Abstract

The blackbody-radiation (BBR) shift remains a leading systematic uncertainty of room-temperature single-ion optical clocks. Accurate evaluation of this shift requires the effective temperature at the ion position, which is difficult because RF heating, low and spatially varying surface emissivity and inaccessible trap surfaces make this quantity difficult to measure directly. Here, we developed a hybrid inversion framework that combines radiance-domain infrared thermography, black-painted high-emissivity reference regions, a finite-element model of a 171Yb+ end-cap ion trap and a reduced-order surrogate for repeated forward calculations and parameter identification. The surrogate reproduced the FE test fields with a full-field mean absolute error (MAE) of 0.02220 mK. The calibrated model predicted the ion-position effective temperature with a combined standard uncertainty of 155.0 mK, corresponding to a temperature-related BBR frequency-shift uncertainty of 1.647 x 10-19 for the 2S1/2-2F7/2 (467 nm) E3 transition. This method provides a reproducible route for reducing and tracing the BBR uncertainty of room-temperature trapped-ion optical clocks.