{DEMAND}: {A}n organic glass scintillator array for direct measurements of (α,n)(α,n) and (p,n)(p,n) reactions in inverse kinematics

G. Christian, B. J. Reed, G. Lotay, G. Bartram, S. Blackmore, D. T. Doherty, A. Katrusiak, J. Henderson, D. A. Hutcheon, A. Lennarz, M. Loria, J. O'Neill, C. O'Shea, C. J. Paxman, C. Ruiz, L. Wagner, M. Williams

physics.ins-det

Abstract

We have developed an array of up to 20 organic glass scintillator detectors, each consisting of a 3 cm×3 cm×3 cm3~\mathrm{cm}\times3~\mathrm{cm}\times3~\mathrm{cm} scintillator cube coupled to a photomultiplier tube. Sensitive to neutrons from a few hundred keV to several tens of MeV, the array was designed for use with recoil separators in inverse-kinematics measurements of astrophysical (α,n)(α,n) and (p,n)(p,n) reactions. It provides a prompt start signal for identifying recoil-neutron coincidences from the recoil time of flight. We characterize eight detectors used in an inverse-kinematics measurement of 22Ne(α,n)25Mg^{22}\mathrm{Ne}(α,n)^{25}\mathrm{Mg}, using source and in-beam data. From the widths of γγ-ray Compton-edge peaks at 311, 341, 477, and 1062 keVee, we find an energy resolution described by FWHM(E)/E=A2/E+B2/E2+C2\mathrm{FWHM}(E)/E=\sqrt{A^2/E+B^2/E^2+C^2}, with A=0.00(7)A=0.00(7), B=0.0618(54)B=0.0618(54), and C=0.135(5)C=0.135(5). The time-difference spectrum of nearest-neighbor γγ-γγ coincidences from a 22Na^{22}\mathrm{Na} source (250--600 keVee) gives a FWHM coincidence timing resolution of 910(40)910(40) ps, corresponding to a single-detector resolution of 640(30)640(30) ps. Mixed neutron-γγ fields from the in-beam reaction and a 241AmBe^{241}\mathrm{AmBe} source show pulse-shape discrimination with an nn-γγ figure of merit >1>1 down to about 100 keVee. Below 100 keVee, the total neutron yield can still be extracted using two-Gaussian fits to energy-gated pulse-shape discrimination spectra. Monte Carlo simulations of the present 22Ne(α,n)25Mg^{22}\mathrm{Ne}(α,n)^{25}\mathrm{Mg} measurement give a neutron singles detection efficiency of 1.56(45)1.56(45)%, with uncertainty dominated by systematic effects in the simulation inputs.

Create a lesson

Metadata

Categories
physics.ins-det, nucl-ex
Comments
Accepted for publication in Nucl. Instrum. and Meth. in Phys. Res. A