The influence of cold wall thickness and background activity on recovery coefficients in PET and SPECT
Abstract
Purpose: The partial volume effect caused by limited spatial resolution consists of signal spill-out from and, in the case of background, spill-in into objects. This can be described by the recovery coefficient (RC), which can be measured in phantoms. With background activity present the RC of an object inside a phantom is influenced by its confining cold wall. This is relevant for the use of RC-values in quantitative imaging, for standard uptake values, and dosimetry. Methods: A theoretical equation for RC as a function of object-to-background concentration ratio (OBR) is formulated and combined with a correction for wall thickness. Novel analytical expressions for spheres and large objects of arbitrary size are derived for the RC of the mean (RCmean). For spheres this expression was compared to previously published experimental data exploring thick (1.5 mm) and thin (0.18 mm) walls. Results: For 10, 37 and 60 mm spheres, 1 mm wall thickness and isotropic spatially invariant Gaussian point spread function with a full width half maximum in the range 2-20 mm, the minimum-maximum values for the change in RCmean times OBR are 0.03-0.19, 0.05-0.06, 0.03-0.04, respectively. Experimental and theoretical differences in RCmean between spheres with thick and thin walls were in good agreement for 10, 15, 38 and 58 mm spheres, but not for the 20 and 30 mm spheres. Conclusion: When using phantom measurements for determining RC-values it is important to consider the cold wall thickness, otherwise significant errors can arise in the presence of background activity.
Create a lesson
Related papers
A novel model-based equation for size dependent mean recovery coefficients for spheres and other shapes
Shared title terminology · Robin de Nijs
MRI Acquisition Parameters Influence Magnetic Resonance Hydrophone Measurements
Shared title terminology · Hunter Harris, Ethan Bryan, Nathan Thyberg et al.
Ventilation Perfusion Functional Difference Images in Lung SPECT: a linear and symmetrical scale as an alternative to the Ventilation Perfusion Ratio
Shared title terminology · Robin de Nijs, Ronan M. G. Berg, Sofie Lindskov Hansen et al.
Influence of Geometry and Tissue Damage on the Mechanical Behaviour of Skin Meshing
Shared title terminology · Masoumeh Razaghi Pey Ghaleh, Denis O'Mahoney, Kevin Moerman
Biomechanical influence of cavity fillers on periwound soft tissue strains in sacral pressure injuries
Shared title terminology · Coraline Staub-Milants, Gr{é}gory Chagnon, No{é}mie Briot et al.
Cerebral Artery Wall Displacement and Strain from 4D MRI
Shared title terminology · Anika Paul, Sjoerd B. Vos, Mostafa Jamshidian et al.