A Low-Cost MR-Compatible Pneumatic Motion Phantom for Magnetic Resonance Imaging
Jonathan Martin, Millie Barron, Charlotte Sappo
Abstract
We describe the detailed design, construction, and testing of a simple, low-cost, open-source, exclusively pneumatic phantom for gated magnetic resonance imaging (MRI) studies of cyclic motion. The overall phantom cost is approximately $200. It is designed to use widely-available materials and an easy-to-program, adaptable micro-controller. Image contrast is generated with air pumped through mineral oil. We demonstrate the utility of the phantom for simulating simple organ motion when developing and investigating image acquisition and reconstruction methods. Phantom motion was independently characterized by video analysis and found to be highly repeatable, with a cycle-to-cycle period variation of 0.6% and a stroke volume variation of 2.3%. The phantom does not generate electromagnetic interference, making it a suitable option for unshielded ultra-low field (ULF) MRI systems. To our knowledge, this work reports the first motion phantom to be imaged on an ULF MRI system
Create a lesson
Related papers
Towards consensus on practical probabilistic evaluation metrics for clinical decision making in radiotherapy
Jenneke I. de Jong, Steven J. M. Habraken, Erik W. Korevaar et al.
Unmyelinated axons alter the biological interpretation of MRI biomarkers of demyelination
Ali Abdollahzadeh, Ricardo Coronado-Leija, Elizabeth Chasen et al.
Impact-Based Analysis Method (IBAM) to evaluate the primary stability of the pedicle screw: A proof of concept
Julie Gabriel, Yvan Petit, Giuseppe Rosi et al.
Geometry-bridge flow matching for blind rigid-motion correction in head cone-beam CT
Sungho Yun, Seungryong Cho
The relationship between junctional zone, smooth muscle tone and intra-uterine pressure
Nicolas R. Chevalier, Emilie Huon, Julie Poujol et al.
A Soft Sponge-Pillar Electrode Platform for Motion-Tolerant Ambulatory EEG and Sleep Monitoring
Chansoo Kim, Weilun Li, Ryan Andersen et al.