A Compact Single-Flange Low-Energy Inverse Photoemission Spectrometer with Elliptical-Mirror Photon Collection

Jong-Am Hong, Min-Jae Maeng, Kyu-Myung Lee, Seunggu Lee, In-Sung Park, Yongsup Park

physics.ins-detarXiv:2611.19503

Abstract

Low-energy inverse photoemission spectroscopy (LEIPS) is limited by weak photon emission, so photon collection optics largely determine its sensitivity. We compare four photon-collection configurations, each addressing a limitation of its predecessor: a detector beside the sample without collection optics (no-mirror), an off-axis parabolic (OAP) mirror delivering a collimated beam to a head-on photomultiplier tube (PMT), external optics coupling the OAP output to a side-on PMT with about twice the quantum efficiency (OAP-external), and an in-vacuum elliptical mirror focusing the emitted light onto the detector (elliptical). Ray-tracing shows that the elliptical mirror delivers 91.9% of the emitted photons to the detector, compared with 63.3% for the OAP mirror and 3.06% without a mirror, while a plane mirror folding the light path by about 90 allows the electron gun, collection optics and detector to fit on a single 8-inch flange. For C60 films, the elliptical configuration gives the highest lowest unoccupied molecular orbital (LUMO) intensity of the four, 271 counts per second per μA of sample current, 1.74 and 3.23 times those of the OAP and OAP-external configurations and about 45 times that of the no-mirror configuration. The gain over the OAP configuration is comparable to the 1.45-fold increase predicted by ray-tracing, but the filter and detector also differed, so the geometric contribution could not be isolated. The OAP-external configuration gave only 0.54 times the OAP signal despite its higher quantum efficiency. High sensitivity therefore requires both a large collection solid angle and a short optical path kept inside the vacuum.

Create a lesson