Absolute Energy Calibration with the Neutron-Activated Liquid-Source System at BaBar's CsI(Tl) Calorimeter
Johannes M. Bauer
Abstract
The electro-magnetic calorimeter at the BaBar detector, part of the asymmetric B Factory at SLAC, measures photons in the energy range from 20 MeV to 8 GeV with good resolution. The calorimeter is calibrated at the low energy end with 6.13 MeV photons obtained from a liquid source system. During the calibration, a fluorine-rich liquid is activated via a neutron generator and pumped past the front of the calorimeter's crystals. Decays that occur in front of the crystals emit photons of well-defined energy, which are detected in the crystals with the regular data acquisition system. The liquid source system adds only very little material in front of the calorimeter, needs nearly no maintenance, and allows operation at the switch of a key with minimal safety hazards. The report describes the system, presents calibration results obtained from its operation since 1999, shows the crystals' loss of light yield due to radiation damage, and shares experiences gained over the years.
Create a lesson
Related papers
Mineral Detection of Neutrinos and Dark Matter 2026 Proceedings
Alexey Elykov, Patrick Stengel, Natsue Abe et al.
Commissioning and Performance of the Time-of-Flight Detector for the T2K Neutrino Oscillation Experiment
C. Alt, L. Amziane, N. Baudis et al.
Design and characterization of large-size monolithic layers of 3D-segmented plastic scintillator
Jon Berisha, Xingyu Zhao, Andrey Boyarintsev et al.
Proposing, never assigning: fair allocation of control-room shifts in experimental particle physics collaborations from ranked wishes under institutional quotas
Kamal Benslama
Terahertz Radar Inversion for Range-Resolved Solids Concentration Profiling in Gas--Solid Flows
Philip Kjaer Jepsen, Albert Monteith, Diana Carolina Guío-Pérez et al.
On the importance of cosmic-ray background in the Atomki anomaly
Hicham Benmansour, Gianluigi Boca, Gianluca Cavoto et al.