The detective Modulation Transfer Function applied to scanned digital radiographic detectors
D. J. Miller, A. Papanestis
Abstract
An extension of the traditional modulation transfer function (MTF), the detective MTF, is proposed for the evaluation of the resolution of X-ray imaging detectors with non-uniform efficiency on the scale of individual detector elements. The dMTF is shown to give a better representation than does the MTF of the improved resolution reported when the readout of the detectors is enhanced.