The instrument response function in air-based scanning tunneling microscopy
P. Fraundorf, J. Tentschert
Abstract
The distinction between point and line resolution in transmission electron microscopy (TEM) arises because an ability to image sub-0.2 nm fringes is a necessary, but not a sufficient, condition for imaging individual atoms. In scanned tip microscopy, as in TEM, empirical data on instrument response should precede assertions about point resolution. In the ``slow scan limit'', time-domain noise and geometry effects decouple, and tip shape can take on the role of a 2-dimensional impulse response function. We indicate here that nuclear track pits can be used to quantitatively measure tip geometry with nanometer-scale resolution in three dimensions, that stationary tip images provide a robust measure of time-domain instabilities, and that when these data are taken before and after imaging an unknown, images with instrument response quantitatively constrained by experiment are possible. Specimen-induced tip effects also become measurable in situ.
Create a lesson
Related papers
Mineral Detection of Neutrinos and Dark Matter 2026 Proceedings
Alexey Elykov, Patrick Stengel, Natsue Abe et al.
Commissioning and Performance of the Time-of-Flight Detector for the T2K Neutrino Oscillation Experiment
C. Alt, L. Amziane, N. Baudis et al.
Design and characterization of large-size monolithic layers of 3D-segmented plastic scintillator
Jon Berisha, Xingyu Zhao, Andrey Boyarintsev et al.
Proposing, never assigning: fair allocation of control-room shifts in experimental particle physics collaborations from ranked wishes under institutional quotas
Kamal Benslama
Terahertz Radar Inversion for Range-Resolved Solids Concentration Profiling in Gas--Solid Flows
Philip Kjaer Jepsen, Albert Monteith, Diana Carolina Guío-Pérez et al.
On the importance of cosmic-ray background in the Atomki anomaly
Hicham Benmansour, Gianluigi Boca, Gianluca Cavoto et al.