X-ray Imaging Using a Hybrid Photon Counting GaAs Pixel Detector
C. Schwarz, M. Campbell, R. Goeppert, E. H. M. Heijne, J. Ludwig, G. Meddeler, B. Mikulec, E. Pernigotti, M. Rogalla, K. Runge, A. Soldner-Rembold, K. M. Smith, W. Snoeys, J. Watt
Abstract
The performance of hybrid GaAs pixel detectors as X-ray imaging sensors were investigated at room temperature. These hybrids consist of 300 mu-m thick GaAs pixel detectors, flip-chip bonded to a CMOS Single Photon Counting Chip (PCC). This chip consists of a matrix of 64 x 64 identical square pixels (170 mu-m x 170 mu-m) and covers a total area of 1.2 cm**2. The electronics in each cell comprises a preamplifier, a discriminator with a 3-bit threshold adjust and a 15-bit counter. The detector is realized by an array of Schottky diodes processed on semi-insulating LEC-GaAs bulk material. An IV-charcteristic and a detector bias voltage scan showed that the detector can be operated with voltages around 200 V. Images of various objects were taken by using a standard X-ray tube for dental diagnostics. The signal to noise ratio (SNR) was also determined. The applications of these imaging systems range from medical applications like digital mammography or dental X-ray diagnostics to non destructive material testing (NDT). Because of the separation of detector and readout chip, different materials can be investigated and compared.
Create a lesson
Related papers
Design and performance of the Fast Beam Condition Monitor for luminosity and background measurement at the CMS Experiment in LHC Run 3
The CMS BRIL Collaboration, Eliana Acurio, Ying An et al.
AlGaN/GaN Hall-Effect Sensor for In-Situ Magnetic Field Monitoring of the HSX Stellarator
Yiming Zhao, Wayne Goodman, Thomas Gallenberger et al.
Measuring and Modelling Lag in Amorphous Silicon Flat-Panel X-ray Detectors
Yiyue Huang, Benjamin Young, Andrew Kingston et al.
Signal formation and induction-gap optimization in a THGEM coupled to a resistive plate anode
Arpan Maity, Luca Moleri, Maryna Borysova et al.
Development and Commissioning of the Cryogenic Target Detectors for the Technical Run of the NUCLEUS Experiment
N. Schermer, H. Abele, G. Angloher et al.
Aliased noise characterization and mitigation in BICEP Array 150, 220 and 270 GHz time-division multiplexed detectors
S. Fatigoni, P. A. R. Ade, Z. Ahmed et al.