CBCT-IQ: A Publicly Available Annotated Cone-Beam CT Dataset for Image Quality Assessment and Benchmarking
Sepideh Hatamikia, Anna Breger, Clemens Karner, Birgit Pohn, Poorya MohammadiNasab, Martin Buschmann, Stephanie Nougaret, Laura Haddad, Ali Abbasian Ardakani, Afshin Mohammadi, Paul Apfaltrer, Wolfgang Birkfellner, Alfred Pohl, Ander Biguri, Gernot Kronreif, Carola-Bibiane Schönlieb, Tess Reynolds
Abstract
Medical image quality plays a critical role in diagnostic accuracy, especially in X-ray-based imaging modalities such as cone-beam computed tomography (CBCT), where image quality must be balanced against radiation dose. While expert visual evaluation remains the clinical standard for image quality evaluation, it is time-consuming, subjective and affected by inter-observer variability, emphasizing the need for reliable quantitative image quality assessment (IQA) methods. However, the development and validation of such IQA methods have been limited by the lack of publicly available CBCT datasets with expert image quality annotations. In this study, we provide the first open-access CBCT IQA dataset containing 1,764 annotated image slices acquired using systematic variations in image acquisition and reconstruction parameters. Three clinical experts graded the overall image quality and a predefined regions of interest (ROI) using a four-level scoring scheme. In addition, we benchmark 26 full reference- and no reference-based IQA measures against expert annotations and introduce an exploratory IQA measure-based ranking capable of distinguishing subtle image quality differences. This dataset introduced a standardized benchmark for future CBCT IQA research and provides a valuable resource for the development and validation of new IQA methods, enabling reproducible research and advancing CBCT IQA.
Create a lesson
Related papers
Cone-beam artifact reduction in Gamma Knife CBCT images using a line-arc-line scan trajectory
Shared title terminology · Alexandra Alain-Beaudoin, Håkan Nordström, Luc Beaulieu et al.
Non-circular scan trajectories for reducing cone-beam artifacts in Gamma Knife CBCT images: a simulation study
Shared title terminology · Alexandra Alain-Beaudoin, Håkan Nordström, Luc Beaulieu et al.
COBRA2026: a large-scale multicenter pelvic cone-beam computed tomography projection dataset
Shared title terminology · Adrian Thummerer, Simon Rit, Florian Kamp et al.
Geometry-bridge flow matching for blind rigid-motion correction in head cone-beam CT
Shared title terminology · Sungho Yun, Seungryong Cho
P^2-Gaussian: Prior-Driven and Projection-Guided Gaussian Splatting for Sparse-View Cone-Beam CT Reconstruction
Shared title terminology · Jinqiu Xia, Jianhui Ma, Bin Ouyang et al.
Bridging cone-beam CT and MRI to stopping power ratio maps: a modality-agnostic Brownian bridge approach to support robust adaptive proton therapy
Shared title terminology · Yuhao Yan, Qisheng He, Minglei Kang et al.