Evaluation of Image Noise from Computed Tomography in an Organ-Specific Anthropomorphic Phantom
Lucas McCullum, Mojtaba Zarei, Andy Tsai, Michael J. Callahan, Peng Sun, Da Zhang
Abstract
Computed tomography (CT) is used ubiquitously for diagnostic imaging, with novel iterative reconstructions such as hybrid iterative reconstruction with partial model-based component (ASiR-V) and deep learning image reconstruction (DLIR) increasingly utilized. In conjunction, CT vendors have developed a noise index (NI) to reduce radiation dose during protocol development. However, it is unclear how these reconstruction algorithms reduce noise as NI changes, particularly across multiple organs. The goal of this study was to utilize an anthropomorphic phantom to evaluate image noise against the prescribed NI across different organs, determine if DLIR reduces image noise compared to ASiR-V, and use this relationship to determine the maximum volume CT dose index (CTDIvol) reduction at equivalent noise levels. Repeat images were acquired on a GE HealthCare Revolution Apex CT scanner across NI values in an anthropomorphic adult-sized phantom from Kyoto Kagaku with ASiR-V and DLIR. Organs-of-interest were contoured using TotalSegmentator in 3D Slicer. The correlation between the CTDIvol and NI was determined. The standard deviation of the difference in the repeat images was used as a noise metric (NM), correlated with NI, used to determine the NI at equivalent NM between ASiR-V and DLIR, then applied to determine the dose reduction and difference in NM. The relationship between NI and the resulting CTDIvol was found to be proportional to a power law of exponent -2.23. The maximum potential absolute dose reduction from using DLIR was found in the liver at 1.3 mGy with an NI of 7.9. Across all organs, the relative dose reduction increased, the NM-to-NI ratio decreased, and the noise reduction increased across NI. The relationship between image noise and the prescribed NI differed between ASiR-V and DLIR and varied by organ, with model-based dose reductions of up to 1.3 mGy in the liver.
Create a lesson
Related papers
High resolution 3D localization of ultrasound sparse sources using a SAGE algorithm
Sophie V. Heymans, François Rottenberg, Tatiana Gelvez-Barrera et al.
Experimental Assessment of Prompt Gamma Time Imaging in Complex Anatomy
Pavel Kavrigin, Alicia Garnier, Adélie André et al.
Ventilation Perfusion Functional Difference Images in Lung SPECT: a linear and symmetrical scale as an alternative to the Ventilation Perfusion Ratio
Robin de Nijs, Ronan M. G. Berg, Sofie Lindskov Hansen et al.
Estimation of Very High Energy Electron (VHEE) beam parameters using an in-air screen
Joel Valerian, Paul Giansiracusa, Vincent Lu et al.
Emission-associated energy deposition and Cherenkov-dose proportionality in megavoltage photon beams
Xianghuan He, Yao Hao, Changyu Sun et al.
Weight-space programming for auditing clinical decision networks
Humza Nusrat, Omar Nusrat